Model Number RT200
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/09/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The rt200 is not sold in the usa but it is similar to a product which is sold in the usa.The 510(k) for that product is k983112.Method: the complaint rt200 adult dual heated evaqua2 breathing circuit was not returned to fisher & paykel healthcare for evaluation.Our investigation is thus based on the information provided by the hospital.Results: the customer experienced condensation during use of a rt200 circuit, and resulted in ventilator to malfunction.Conclusion: without the complaint device, we are unable to determine what may have caused the condensation reported.Condensate in the humidification system, although not preferred, is an expected side effect of heated pass-over humidification systems in many conditions, and may vary between light misting to water droplets that form on the wall of cool breathing circuit tubing.The amount of condensate in the ventilation system is influenced by a number of multiple setup and environmental factors.Our user instructions that accompany the rt200 adult breathing circuit state the following: set appropriate ventilator alarms.Check that the heater wire is evenly distributed along the circuit and not bunched or kinked.Do not stretch or milk the tubing.When mounting a humidifier adjacent to a patient ensure that the humidifier is always positioned lower than the patient.Check breathing circuits for condensation every 6 hours and drain if required.
 
Event Description
A healthcare facility in (b)(6) reported that there was excessive condensate in the rt200 adult single use breathing circuit.The condensate then led to the ventilator to malfunction.There was no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT DUAL HEATED BREATHING CIRCUIT
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key8924029
MDR Text Key176056246
Report Number9611451-2019-00824
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT200
Device Catalogue NumberRT200
Device Lot Number2100714157
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/25/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-