Model Number 20E
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Patient Involvement (2645)
Event Date 07/02/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the device's therapy pcb assembly.After completing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.Physio-control further evaluated the removed therapy pcb assembly and did not observe any issues with this part.
 
Event Description
The customer contacted physio-control to report a non-critical issue with their device.There was no report of patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that the device was delivering monophasic energy.As a result, the wrong defibrillation therapy may be delivered to a patient, if it was needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20E DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key8925856
MDR Text Key155373136
Report Number0003015876-2019-01407
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873813829
UDI-Public00883873813829
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20E
Device Catalogue Number99507-000016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2019
Date Manufacturer Received07/31/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-