This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on august 23, 2019.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Method code #1: 10 - testing of actual/suspected device.Method code #2: 11 - testing of device from same lot/batch retained by manufacturer.Method code #3: 3331 - analysis of production records.Results code: 3259 - improper physical structure.Conclusions code: 4307 - cause traced to component failure.There were 6 samples returned for evaluation and found the arterial thermistor to be functioning on all 6.A representative retention sample from the same product/lot number combination was tested and found to be functioning properly.The root cause of the issue has been determined to be a high insertion force during assembly of the thermistor into the oxygenator module coupled with exposure to extreme temperature fluctuations.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|