MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Use of Incorrect Control/Treatment Settings (1126)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6).Date of report: 27aug2019.The manufacturer¿s international service technician confirmed the reported flow accuracy issue.The manufacturer¿s international service technician replaced the flow sensor assembly to address the reported problem.The unit was checked overall, run in tested, cleaned and functionally tested and no abnormality was confirmed.The gas delivery system (gds) was return for analysis.An investigation was performed and the oxygen flow sensor u1 on the customer returned gds had an offset that caused the air flow accuracy test to fail at the 10lpm test point and the oxygen flow accuracy test to fail at the 0 lpm test point.Replacing u1 resolved the problem, both air and o2 flow accuracy tests passed.This mdr has been reassessed as reportable after a request from the fda on march 1, 2019 to review complaints from 29 nov 2017 to 29 nov 2018.As this has been reassessed, it will appear to be a late mdr.
 
Event Description
During periodic maintenance, the manufacturer¿s international service technician noted that when the oxygen flow setting was zero slpm, the device indicated the value of 1.8 slpm.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key8935754
MDR Text Key176840088
Report Number2031642-2019-07348
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2018
Was the Report Sent to FDA? No
Date Manufacturer Received04/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-