PHYSIO-CONTROL, INC. - 3015876 LUCAS¿ 2 CPR CHEST COMPRESSION SYSTEM; COMPRESSOR, CARDIAC, EXTERNAL
|
Back to Search Results |
|
Catalog Number 99576-000033 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
Death (1802)
|
Event Date 08/05/2019 |
Event Type
Death
|
Manufacturer Narrative
|
(b)(4).Stryker evaluated the customer's device and was unable to duplicate the reported issue.After performing an unrelated repair, proper device operation was observed.The device was subsequently returned to the customer for use.The cause of the reported issue could not be determined.Stryker completed a clinical review and determined that the device use may have contributed to the patient outcome, as the interruption in cpr was one minute.The users did not follow the instructions for use that state if there are interruptions, to remove the device and start manual chest compressions.By following the instructions for use the interruption in cpr is likely to be short, not, as in this case, as long as one minute.
|
|
Event Description
|
The customer contacted stryker to report that their device would not work during a patient event.The device indicated the battery was fully charged, however the device did not operate.The device's suction cup was placed on the patient, but the device would not operate when the continuous and 30:2 buttons were pressed.Eventually, the device was removed from the patient and manual cpr was provided.When the device would not operate and it was removed from the patient, it was plugged into the power supply, to charge the battery for 10 minutes.The device subsequently operated correctly.The interruption in cpr was estimated to be approximately 1 minute.It was delayed due to decision making around transporting without a lucas, while it was being charged.The patient did not survive the reported event.
|
|
Event Description
|
The customer contacted stryker to report that their device would not work during a patient event.The device indicated the battery was fully charged, however the device did not operate.The device's suction cup was placed on the patient, but the device would not operate when the continuous and 30:2 buttons were pressed.Eventually, the device was removed from the patient and manual cpr was provided.When the device would not operate and it was removed from the patient, it was plugged into the power supply, to charge the battery for 10 minutes.The device subsequently operated correctly.The interruption in cpr was estimated to be approximately 1 minute.It was delayed due to decision making around transporting without a lucas, while it was being charged.The patient did not survive the reported event.
|
|
Manufacturer Narrative
|
Section h6, results code of the initial medwatch report indicates no device problem found.Section h6, results code of the initial medwatch report should indicate power source problem identified.Section h6, conclusion code on of the initial medwatch report indicates cause not established.Section h6, conclusion code of the initial medwatch report should indicate shelf life/expiration date exceeded.Section h10 and/or h11 on of the initial medwatch report indicates stryker evaluated the customer's device and was unable to duplicate the reported issue.After performing an unrelated repair,proper device operation was observed.The device was subsequently returned to the customer for use.The cause of the reported issue could not be determined.Stryker completed a clinical review and determined that the device use may have contributed to the patient outcome, as the interruption in cpr was one minute.The users did not follow the instructions for use that state if there are interruptions, to remove the device and start manual chest compressions.By following the instructions for use the interruption in cpr is likely to be short, not, as in this case, as long as one minute.Section h10 and/or h11 of the initial medwatch report should indicate stryker completed a clinical review and determined that the device use may have contributed to the patient outcome, as the interruption in cpr was one minute.The users did not follow the instructions for use that state if there are interruptions, to remove the device and start manual chest compressions.By following the instructions for use the interruption in cpr is likely to be short, not, as in this case, as long as one minute.Stryker evaluated the customer's device and was unable to duplicate the reported issue.After performing an unrelated repair,proper device operation was observed.The device was subsequently returned to the customer for use.It was noted that the batteries used with the device were manufactured in 2015.Stryker recommends that batteries be replaced approximately every two years.The likely cause of the reported issue was determined to be due to the battery.
|
|
Manufacturer Narrative
|
It was noted that the batteries used with the device were manufactured in 2015 and should be replaced.Stryker recommends that batteries be replaced approximately every three years.The likely cause of the reported issue was determined to be due to user error, as the customer had not replaced the battery after the recommended 3 year life.
|
|
Event Description
|
The customer contacted stryker to report that their device would not work during a patient event.The device indicated the battery was fully charged, however the device did not operate.The device's suction cup was placed on the patient, but the device would not operate when the continuous and 30:2 buttons were pressed.Eventually, the device was removed from the patient and manual cpr was provided.When the device would not operate and it was removed from the patient, it was plugged into the power supply, to charge the battery for 10 minutes.The device subsequently operated correctly.The interruption in cpr was estimated to be approximately 1 minute.It was delayed due to decision making around transporting without a lucas, while it was being charged.The patient did not survive the reported event.
|
|
Search Alerts/Recalls
|
|
|