Model Number 24970A |
Device Problem
Computer Operating System Problem (2898)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/28/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the programmer froze during a programming session.The user had attempted to enter the threshold test results in the patient information screen, but the screen would not respond to input.The programmer remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product analysis : analysis revealed that the complaint description matches with log dated on 28-aug.When user is viewing the report, they pushed application to background and foreground multiple times, which causes sockets to close and not resume.This is one of background/foreground issue and is documented in rpi (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|