Investigation visual inspection no significant deformations or damage of the valve was detected during the visual inspection.Permeability test a permeability test has shown that the progav 2.0 valve is permeable.Adjustment test the progav 2.0 valve was tested and is adjustable to all specified pressures.Braking force and brake function test the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Computer controlled test to investigate the claim of over/under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The progav 2.0 valve is operating within acceptable tolerances.Result first, we performed a visual inspection of the progav 2.0 valve.No significant deformations or damage of the valve was detected during the visual inspection.Next, we tested the permeability, adjustability, and opening pressure of the valve, as well as the brake functionality and brake force.The progav 2.0 valve operates as expected and met all specifications.Finally, we have dismantled the valve.Inside the valve, we have found a build-up of substances (likely protein).Based on our investigation, we are unable to substantiate the claim of non-adjustability.At the time of our investigation, the progav 2.0 could be adjusted to all specified settings and the opening pressure is in the acceptable tolerances.This is likely due to the deposits observed inside the valves.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
|