Model Number 1458Q/86 |
Device Problem
Difficult to Advance (2920)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/27/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that during the implant procedure, the guidewires could not advance into the lead or were entrapped to the leads.All three of the leads were scrapped and replaced.The patient was stable.
|
|
Manufacturer Narrative
|
Additional information: correction: the reported event of "trapped wire" was confirmed.The lead was tested with a guidewire and restriction was noted at the middle region.Visual examination found a fibrous foreign material in the inner coil.The cause of the event was isolated to the discovered foreign material.
|
|
Event Description
|
Related manufacturer reference number: 2938836-2019-13284, 2938836-2019-13287.
|
|