• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P6
Device Problem Optical Problem (3001)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to the service center.A visual inspection on the received condition was performed and found a portion of the bending section skeleton protruding out from the bending section cover near the insertion tube side.The bending section damage is approximately 70mm from the distal end side, which is causing a leak on the bending section cover.The bending section cover was removed in order to reveal the extent of the damage.Upon removal of cover, it was confirmed that the bending section skeleton is fully separated producing sharp edges near the insertion tube side.The bending section support pins are still intact and not lifted.The scope was previously refurbished (28f) on june 20th 2019.Further finding¿s include; excessive broken fibers (black dots) throughout the image, a c-body leak and corroded bending section.Based on the evaluation of the returned device, the most likely cause of the broken fibers and bending section skeleton breaking is due to excessive force and/or stress, attributed to user mishandling.The instructions for use contains the following warnings to prevent equipment damage and patient injury: "do not twist or bend the bending section with your hands.Equipment damage may result; do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damage.If any of the following conditions occur during an examination, immediately stop the examination and withdraw the endoscope from the patient.If the up/down angulation control lever does not move.If the angulation control mechanism is not functioning properly.Visually inspect the bending section for no metallic parts protruding from the bending section.Visually inspect the bending section for bends, twist, or other irregularities while the bending section remains straight.Visually inspect the bending section for abnormal bending shape, or other irregularities.Continued use of the endoscope under these conditions could result in patient injury, bleeding, and/or perforation.".
 
Event Description
The user facility reported that the device in question has broken fibers (ig fiber breakage).During service activities the bending skeleton was noted to be broken with metal protruding.There was no patient involvement or adverse outcome reported as a result of the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key8985525
MDR Text Key200446621
Report Number8010047-2019-03248
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2019
Was the Report Sent to FDA? No
Date Manufacturer Received08/13/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-