Model Number 3662
Device Problems Premature End-of-Life Indicator (1480); No Apparent Adverse Event (3189)
Patient Problem Inadequate Pain Relief (2388)
Event Date 08/19/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that patient received 'replace generator soon' message.At time of visit with rep the message was cleared, however it then returned.Surgical intervention may be taken at a later date to address the issue.
 
Manufacturer Narrative
The device was not returned for analysis; however, logs and/or assessment report were assessed and indicate that the battery voltage level of the device is within specifications to trigger the eri indicator.Based on the information received, the cause of the reported incident is consistent with the battery nearing normal end of life.
 
Event Description
It was reported that the old ipg was explanted and a new ipg implanted.No reported complications.Therapy reportedly was restored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 7 ELITE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9008507
MDR Text Key157799760
Report Number1627487-2019-10267
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020222
UDI-Public05415067020222
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/03/2020
Device Model Number3662
Device Lot Number6242832
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/04/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight113
-
-