The device was not returned for evaluation.
A potential failure mode could be '8.
2 bio-incompatible- effect 2: reaction¿ with a potential root cause of '8.
2.
2 materials that are contacting the patient's intact skin are not biocompatible.
¿ the lot number is unknown; therefore, the device history record could not be reviewed.
The instructions for use were found adequate and state the following: "there are no known contraindications for the use of a noninvasive thermoregulatory system.
Do not place the neonatal arctic gel pad on skin that has signs of ulcerations, burns, hives or rash.
Do not remove the fabric release liner of the neonatal arctic gel pad and expose the hydrogel.
While there re no known allergies to hydrogel materials, caution should be exercised with any patient with a history of ski allergies or sensitivities.
".
|