Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. .
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: added: h6.Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Per internal procedures, the event information was reviewed.No investigational inputs were received.For this investigation, no immediate action was required as no alleged deficiency with the device was identified.No device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot number combination.A search of the depuy non-conformance (nc) quality system found no nc¿s associated with this product/lot combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.Investigational inputs were requested as indicated per internal procedures for this failure mode, however, no information was provided to depuy without the physical complaint sample associated with this report, it was not possible to determine if the device failed to meet specifications at the time it was released for distribution.The device associated with this event was used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No evidence was found indicating product error was a contributing factor.The need for corrective action was not indicated.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot :j01j59.Device history review: a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.
|