Product analysis #(b)(4): analysis information -- 2019-09-10 13:35:00 cst pli# 10 product id# 8780 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection, patency testing, and pressure testing.Visual inspection identified a broken seal in the sterile package.The device history report (dhr) was reviewed and there were {no} anomalies in the manufacturing processes of the device.Analysis determined the inner tray of the sterile packaging did not properly hold the components in place.The pump was returned, and analysis found the inner tray did no properly hold the components in place and the sterile seal was compromised.If information is provided in the future, a supplemental report will be issued.
|