DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT20649-1
Device Problem Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/14/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a unexpected receiver shut-down occurred. No product or data was provided for evaluation. Confirmation of the allegation and a root cause could not be determined. No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The product was evaluated. The device was visually inspected and it passed. The receiver failed to charge and boot due to no power but will turn on with a known good battery after receiver case is open. The receiver download passed by using a known good battery. The log was downloaded and reviewed finding firmware errors. The receiver case was opened and the internal visual inspection failed due to finding battery damage. The allegation was confirmed. The probable cause is damaged battery. No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key9015952
MDR Text Key160734948
Report Number3004753838-2019-65671
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/12/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/12/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMT20649-1
Device Catalogue NumberSTR-GL-004
Device Lot Number5248899
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/09/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-