RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Device Alarm System (1012); Incorrect, Inadequate or Imprecise Result or Readings (1535); Component Missing (2306)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2018.Date of report: 16sep2019.The manufacturer¿s international service technician confirmed the reported pressure issues.The manufacturer¿s international service technician replaced the data acquisition (da) pcba to address the reported problem, replaced the power switch overlay and replaced the fan filter housing.Failure analysis of the returned part indicates: no root cause identified the unit was checked overall, run in tested, cleaned and functionally tested and no abnormality was confirmed.This mdr has been reassessed as reportable after a request from the fda on march 1, 2019 to review complaints from 29 nov 2017 to 29 nov 2018.As this has been reassessed, it will appear to be a late mdr.
|
|
Event Description
|
During periodic maintenance, the manufacturer¿s international service technician reported that the pressure accuracy test was out of specifications, the led alarm failed, and the fan filter cover was missing.The customer reported there was no patient involvement.
|
|