Catalog Number 955542
Device Problem Protective Measures Problem (3015)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/22/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during continuous renal replacement therapy (crrt) of a patient using a prismaflex control unit, communication error alarm code 2 was generated causing the prismaflex control unit to enter a patient safe state.The treatment was discontinued, and the patient was disconnected without returning the extracorporeal blood in the circuit.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was not available for evaluation.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - LUND MONITORS
lund skane lan
MDR Report Key9078681
MDR Text Key183131532
Report Number9616026-2019-00055
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K131516
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955542
Was Device Available for Evaluation? No
Date Manufacturer Received10/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PRISMAFLEX SET.
-
-