Catalog Number 955542 |
Device Problem
Protective Measures Problem (3015)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during continuous renal replacement therapy (crrt) of a patient using a prismaflex control unit, communication error alarm code 2 was generated causing the prismaflex control unit to enter a patient safe state.The treatment was discontinued, and the patient was disconnected without returning the extracorporeal blood in the circuit.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The device was not available for evaluation.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
|
|