Model Number 16-02-80
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 08/19/2019
Event Type  Malfunction  
Manufacturer Narrative

There was no patient involvement. The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa. (510(k) number: k052601). Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A livanova field service representative in charge found out that the patient pump 2 was stuck. The part was replaced and the issue solved. Subsequent functional verification testing was completed without further issues and the unit was returned to service. Pictures of the affected parts has been provided as evidence. The analysis of the pictures revealed that the bearing damage is visible and this had led to the reported error code outbreak. A review of the dhr could not identify any deviations or non conformities relevant to the issue.

 
Event Description

Livanova (b)(4) received a report that a heater-cooler system 3t device displayed an error code on the patient circuit 2 during maintenance. There was no patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 89309
GM 89309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 89309
GM 89309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9080318
MDR Text Key159106182
Report Number9611109-2019-00726
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeTW
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Remedial Action Other
Type of Report Initial
Report Date 09/17/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received09/18/2019
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-80
Was Device Available For Evaluation? No Answer Provided
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received08/19/2019
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured08/29/2016
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-