• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Model Number TSH
Device Problem Low Test Results (2458)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/16/2019
Event Type  malfunction  
Manufacturer Narrative
The serial number for the cobas e 801 used at the investigation site was (b)(4).The tsh reagent lot used on this cobas e 801 was 386646 with an expiration date of 30-may-2020.The ft4 iii reagent lot used on this cobas e 801 was 380330 with an expiration date of 31-dec-2019.The ft3 iii reagent lot used on this cobas e 801 was 348359 with an expiration date of 31-oct-2019.The investigation is ongoing.This event occurred in (b)(6).
 
Event Description
The initial reporter complained of questionable thyroid results for 1 patient sample on a cobas 8000 e 801 module.The patient sample was submitted for investigation where discrepant results were identified for elecsys tsh assay, elecsys ft4 iii assay, and elecsys ft3 iii between the customer's e 801 module and an e 801 module used at the investigation site compared to the siemens centaur method used at the investigation site.It was asked but not known if the results from the customer site were reported outside of the laboratory.This medwatch will cover tsh.Refer to medwatch with patient identifier (b)(6) for information on the ft4 iii results, and medwatch with patient identifier (b)(6) for information on the ft3 iii results.The customer's cobas e 801 serial number (b)(4).
 
Manufacturer Narrative
All values, generated with the analyzers from roche diagnostics and the analyzer from siemens, are within the normal reference ranges of the respective assays.The observed differences in tsh values generated with the roche assay and siemens centaur assay are most likely caused by differences in the overall setups of the assays, the antibodies used, differences in reference materials, and the differences in the standard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH ASSAY
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key9083426
MDR Text Key219199912
Report Number1823260-2019-03376
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 10/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model NumberTSH
Device Catalogue Number07028091190
Device Lot Number365417
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-