The investigation has determined that a higher than expected result was obtained from a vitros performance verifier (pv) quality control (qc) fluid using vitros chemistry products k+ slides on a vitros 350 chemistry system.The assignable cause of the higher than expected vitros k+ quality control result is most likely a suboptimal calibration.The slope and intercept values of the initial calibration of vitros k+ reagent lot 4102-1016-9431 were abnormally high compared to the release values.After the reagent was calibrated a second time on the same day, the slope and intercept values of the recalibration were comparable to the release values.The quality control results, and a patient sample result tested after the second calibration event were deemed acceptable.The cause of the suboptimal calibration is unknown.Based on historical quality control results as well as qc results after the second calibration event on (b)(6) 2019, a vitros k+ reagent lot performance issue is not a likely contributor to the event.In addition, a vitros cl- calibration was determined to be suboptimal at the time of the atypical vitros k+ calibration.After both assays were calibrated a second time, optimal calibrations were obtained for each assay.Furthermore, continual tracking and trending of complaint data has not identified any signals to suggest there is a systemic quality issue with vitros k+ reagent lot 4102-1016-9431.Within run precision testing was not performed on the vitros 350 chemistry system.However, there was no evidence to suggest that the vitros 350 chemistry system was a contributing factor of the event.The higher than expected result was obtained from a quality control fluid and no results were reported from the laboratory.No patient sample results were reported from the laboratory.Ortho has not been made aware of any allegation of patient harm as a result of this event.
|