It was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup and bhr head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the head and cup.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.The documentation shows the acetabular cup met manufacturing specifications at the time of production.Non-conformance was identified for the orchid casting certificate of conformity for the femoral head as part of the dhr could not be located.However, the machining and sterilisation and all supporting documents confirm that the device will have met all conforming standards at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.The reported elevated cocr levels and altr may be consistent with an adverse reaction to metal debris, but this cannot be confirmed based on the limited information provided.Without supporting radiographic images, lab/pathology results, and/or the analysis of the explanted components, the root cause of the reported clinical symptoms cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the revision and associated postop pain cannot be determined, and there is no report of the patient¿s current condition, or whether the reported clinical symptoms persist.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be re-opened.No preventative or corrective action has been initiated as a result of this investigation.
|