Model Number C6TR01
Device Problems Unable to Obtain Readings (1516); Loss of Data (2903); Reset Problem (3019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/29/2019
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: 419678 lead, implanted: (b)(6) 2015. If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that on the interrogation report for the cardiac resynchronization therapy pacemaker (crt-p), the patient data and pacing percentage counters were indicated as question marks and the histogram report showed invalid data. It was also reported that bit flip errors and power on resets were identified. Additional information reported that the patient has an abandoned plutonium powered implantable pulse generator (ipg) system which remains in their body. The crt-p remains in use. No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis. However, performance data collected from the device was received and analyzed. Analysis of the device memory indicated an issue with missing/invalid diagnostic data. If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVIVA CRT-P
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9112748
MDR Text Key159925247
Report Number9614453-2019-03209
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00643169351905
UDI-Public00643169351905
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2019
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date02/28/2016
Device Model NumberC6TR01
Device Catalogue NumberC6TR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 09/24/2019 Patient Sequence Number: 1
Treatment
5024M-58 LEAD, 4269 LEAD
-
-