Catalog Number JC-05400-B |
Device Problem
Mechanics Altered (2984)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/27/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
|
Event Description
|
It was reported that while inserting the peridural catheter, the needle included in the kit was not sealed: air vacuum during aspiration test after insertion of catheter.Opened a new kit and no problem with the new needle.
|
|
Event Description
|
It was reported that while inserting the peridural catheter, the needle included in the kit was not sealed: air vacuum during aspiration test after insertion of catheter.Opened a new kit and no problem with the new needle.
|
|
Manufacturer Narrative
|
Qn#(b)(4).A device history record review was performed on the lor syringe with no relevant findings.A design history review was performed for part # kz-05501-002 as a part of this complaint investigation.Per eco-051699 (released 03-dec-2018), supplier (preox) made the following changes: kz-05501-002 luer plastic lor syringe: a.Changed plunger material from profax 535 to profax 531.B.Changed to new plunger tool.C.Changed to new mold for blue stopper.D.Changed molding location for the plunger and the blue stopper as follows.Plunger: from fleimaplastic in germany to gpe, germany.Blue stopper: from et, germany to psilkon, germany.These effective changes did impact product design and material.A corrective action is not required at this time as the root cause for this complaint investigation could not be determined based upon the information provided and without the actual sample.Complaint verification testing could not be performed as no sample was provided for analysis.A device history record review was performed on the lor syringe with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of the lor syringe leaking could not be determined based upon the information provided and without the sample.
|
|
Search Alerts/Recalls
|