Investigation evaluation: on-going.When additional information becomes available a follow up report will be submitted.Re-submission of this mdr due to incorrect initial mdr number of 3004721439-2019-00236.This mdr was submitted on 05 sept 2019 and acknowledgements passed, however, the correct mdr number for this complaint is 3004721439-2019-00233.
|
Investigation visual inspection no significant deformations or damage of the valve were detected during the visual inspection.Permeability test a permeability test has indicated that both valves have blackages.Adjustment test the progav 2.0 valve was tested and is adjustable to all specified pressures.Braking force and brake function test the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Results it should be noted that the shunt system was received dry (i.E.Not submersed in liquid as recommended).The investigation of dry items is not significant due to the affect dry depositis of kiquor and blood can have on product performance.In spite of this, we have investigation the system to the vest of our abilities.First, we performed a visual inspection of the progav 2.0 shunt system.No significant deformation or damage of the valves were detected during the visual inspection.Next, we tested the permeability and opening pressure of the valves.Both valves were shown to be non-permeable, indicating blockage.Additionally, we tested the adjustability as well as the brake functionality and brake force of the progav 2.0 valve.The valve operated as expected and met all specifications.Finally, we have dismantled the valves.Inside both valves we have found significant build-up of substances (likely protein).Based on our investigation, we confirm the presence of occlusion in the system at the time of investigation.This is likely due to the deposits observed inside the valves.As described in our literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
|