• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AVANCE CS2; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AVANCE CS2; ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1009-9050-000
Device Problem Defective Alarm (1014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/21/2019
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system but did not confirm the reported issue.The alarm speakers were replaced.Customer contact reports no patient information available.Date received by manufacturer: this mdr malfunction event was previously eligible for the voluntary malfunction summary reporting (vmsr) program.As of 9/16/2019 notification from fda, this product code is no longer eligible for vmsr.According to the notification, ge healthcare must submit individual reports within 30 calendar days of receiving the notification.Therefore, this event is being reported as an individual mdr report due 10/16/2019.
 
Event Description
The hospital reported a speaker not detected error which could result in the loss of critical alarms.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANCE CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
MDR Report Key9123021
MDR Text Key177265902
Report Number2112667-2019-00631
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier00840682102322
UDI-Public010084068210232221
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1009-9050-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-