DEPUY SYNTHES PRODUCTS LLC COLIBRI II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.101 |
Device Problem
Intermittent Energy Output (4025)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/13/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi: (b)(4).The actual device was returned for evaluation.During evaluation of the device, it was determined that the device would not run - motor damaged.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to premature wear.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that the small battery drive device was working on and off when the trigger was pressed and had a significant lack of power.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available has been disclosed.If additional information were to become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|