Model Number URF-V2
Device Problems Break (1069); Material Protrusion/Extrusion (2979)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to the service center for evaluation. A visual inspection was performed on the received device and noted the bending section skeleton was exposed from the torn bending section cover causing it to fail the leak test. The bending section cover was then removed and noted the bending section skeleton tab was fully broken/detached at the insertion tube area. Additionally, there was no significant evidence of sharp areas noted with the broken/detached skeleton. Based on the evaluation the cause of the broken skeleton is due to excessive stress and force. In the instruction manual it states that ¿do not twist or bend the bending section with your hands. Equipment damage may result; do not insert the insertion tube with excessive force into the ureter or calix. The bending section maybe damage.
 
Event Description
The service center was informed during the preparation for use there were exposed wires noted on the bending section the scope. There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameURETERO-RENO VIDEOSCOPE
Type of DeviceURETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key9130654
MDR Text Key219551187
Report Number8010047-2019-03407
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial
Report Date 09/27/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberURF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2019
Is the Reporter a Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/02/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-