ST. JUDE MEDICAL - NEUROMODULATION EON RECHARGEABLE IPG, 16-CHANNEL SCS IPG Back to Search Results
Model Number 3716
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 09/10/2011
Event Type  malfunction  
Manufacturer Narrative
This device is included in the fsca 1627487/07/26/2012/001-c. Date of event is estimated. Attempts were made to obtain complete patient and event information. Further information was not provided. The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Reference manufacturer numbers: 1627487-2019-10835. The investigation for capa 118058 associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns. The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameEON RECHARGEABLE IPG, 16-CHANNEL
Type of DeviceSCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9131215
MDR Text Key160450607
Report Number1627487-2019-10834
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Remedial Action Recall
Type of Report Initial
Report Date 09/27/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date05/28/2014
Device Model Number3716
Device Catalogue Number3716
Device Lot Number3699664
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/10/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/07/26/2012/001-C

-
-