• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC COLIBRI II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC COLIBRI II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.101
Device Problem Unexpected Shutdown (4019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Reporter¿s complete mailing address was not provided.The actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the small battery drive device stopped working.It was not reported if there were any delays in the surgical procedure or whether a spare device was available for use.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown but was noted to have occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was not confirmed.However during evaluation, it was noted that that the device made unexpected noise due to faulty parts.The assignable root cause was traced to component failure.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLIBRI II
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9134648
MDR Text Key160540878
Report Number8030965-2019-68730
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2019
Date Manufacturer Received12/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-