ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3562
Device Problems Premature Discharge of Battery (1057); Failure to Interrogate (1332)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/16/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that the patient presented in clinic for a follow-up when it was noted that the device failed to be interrogated.The patient was sent to the hospital and was on temporary pacing.The device was explanted and replaced successfully.Patient¿s international normalized ratio (inr) was high prior to the procedure.Patient condition was stable after the procedure.
 
Manufacturer Narrative
The device was received from the field with no telemetry and battery having reached end of service level.Attempts to interrogate the device failed due to lack of telemetry caused by low battery voltage.Following x-ray evaluation, which showed no anomaly, the pacer was cut-open and connected to an external power source which indicated high drain current.The device was tested further using thermal imaging which narrowed down to the capacitor as the source of high leakage and depletion of the battery consistent with the communication anomaly observed on the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key9136737
MDR Text Key160772915
Report Number2938836-2019-14184
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberP000046258
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/27/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age48 YR
Patient Weight63
-
-