MAQUET CARDIOPULMONARY GMBH HEATER COOLER UNIT; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 701044054 - HCU 40
Device Problem Insufficient Cooling (1130)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/10/2019
Event Type  malfunction  
Manufacturer Narrative
A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
Ice loss during the operation.Warm water runs back into the tank and makes the ice melt.(b)(4).
 
Manufacturer Narrative
According to the service report dated on 2019-09-10 the field service technician has replaced the following components: valve kit (material#: 70107.1778), vacuum valves (material#: 70106.4680) and shunt valve (material#: 70107.1568).This case has already been investigated in complaint: (b)(4): according to the investigation report lce04268 dated on 2019-09-06 the valves are defective.They do not close completely, it can happen that warm water flows into the tank and the desired amount of ice cannot be built.Due to the visuell investigation deposits on the o-rings was found.The deposits on the o-rings could most likely have led to the leakage of the valve.Thus the reported failure could be confirmed.The hcu 40 risk analysis version v17 (dms#: 2023635) was reviewed and the failure is assessed in section: risk id: h9.2.1.18 possible causes: too low amount of ice building: sensor failures, use of water out of spec.(conductivity).Software error, defective compressor or coolant circuit, software error.Since the incident is below the accepted rate, no remedial action is necessary.The ocurrance rate related to the reported issue is currently being monitored as part of maquet cardiopumonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER COOLER UNIT
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key9138529
MDR Text Key195103833
Report Number8010762-2019-00309
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K130300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Type of Report Initial,Followup
Report Date 10/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number701044054 - HCU 40
Device Catalogue Number701044054
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-