Model Number 210080 |
Device Problems
Material Frayed (1262); Material Integrity Problem (2978)
|
Patient Problem
Tissue Damage (2104)
|
Event Date 09/26/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
As part of the normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
|
Event Description
|
Splintering in the black leg holder has occurred and gave employee in cpd splinters (carbon fiber) that has caused concern as to the structural integrity of the device and if further splintering could occur in the operating room and affect the patient.Would like all carbon pieces replaced (creating individual complaints per piece).Case type: (b)(4).Patient was under anesthesia.
|
|
Manufacturer Narrative
|
Splintering in the black leg holder has occurred and gave employee in cpd splinters (carbon fiber) that has caused concern as to the structural integrity of the device and if further splintering could occur in the or and affect the patient.Would like all carbon pieces replaced (creating individual complaints per piece).Case type: tka patient was under anesthesia.Product evaluation and results: visual inspection: confirms carbon fiber splitting on the boot assembly.Product history review: 1.Review of the device history records indicate 3 devices were manufactured and accepted into final stock on 08-01-2017 with no reported discrepancies.2.Review of the device history records indicate 14 devices were manufactured and accepted into final stock on 12-02-2017 with no reported discrepancies.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 210080, lot 20174308202 shows 1 additional complaints related to the failure in this investigation.Conclusions: the event was confirmed.
|
|
Event Description
|
Splintering in the black leg holder has occurred and gave employee in cpd splinters (carbon fiber) that has caused concern as to the structural integrity of the device and if further splintering could occur in the or and affect the patient.Would like all carbon pieces replaced (creating individual complaints per piece).Case type: tka patient was under anesthesia.
|
|
Event Description
|
Splintering in the black leg holder has occurred and gave employee in cpd splinters (carbon fiber) that has caused concern as to the structural integrity of the device and if further splintering could occur in the or and affect the patient.Would like all carbon pieces replaced (creating individual complaints per piece).Case type: tka.Patient was under anesthesia.
|
|
Manufacturer Narrative
|
"while reviewing the complaint record associated with this report, it was discovered that this event was incorrectly filed as a serious injury.We confirmed that this event did not cause or contribute to any life-threatening condition or permanent impairment to the patient and did not require any medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure.Additionally, a review of complaint history records from january 1, 2016 until today, confirmed that there have been no reports of death or serious injury associated with similar events for this device family.Finally, the device¿s associated risk documentation confirmed that the highest potential severity of harm for this hazardous situation is a s2.Therefore, we will no longer be filing reports for this event type.
|
|