• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH MIS CANN POLY SCW 6X50MM,TI; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH MIS CANN POLY SCW 6X50MM,TI; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 186715650
Device Problem Loose or Intermittent Connection (1371)
Patient Problems Unspecified Infection (1930); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported on an unknown date, patient was in an accident in (b)(6) and taken to a (b)(6) trauma center for a t12 compression fracture.Patient was instrumented percutaneously from t10-l2 using dss viper system last (b)(6) 2019.Upon return to the us, patient was bending and heard/felt a pop.He came to the er and images revealed the hardware had disassembled.Several set screws were displaced and the rod had come out of the screw saddle.During the revision surgery, on (b)(6) 2019, it was found that all of the set screws were loose and some were sitting extremely proud in the screw saddle.Hardware was removed, replaced, and sufficiently locked.There was a hardware failure and infection.There was a patient consequence.Procedure outcome is unknown.Concomitant device reported: viper2 straight rod-160mm (part# 186789160, lot# unknown, quantity 1), viper2 straight rod-180mm (part# 186789180, lot# unknown, quantity 1).This complaint involves sixteen (16) devices.
 
Manufacturer Narrative
Product complaint # (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The screw features signs of use such as surface markings and wear to the drive feature of the screw.However, the screw featured no signs of loosening.No postoperative x-rays were provided to confirm the pedicle screw loosening.The loosening cannot be replicated.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.The root cause of the poly screw loosening postoperatively cannot be determined from the sample and information provided.A potential root cause cannot be determined since the report of the screw loosening could not be confirmed or replicated.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIS CANN POLY SCW 6X50MM,TI
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9153327
MDR Text Key167952627
Report Number1526439-2019-52202
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034122366
UDI-Public(01)10705034122366
Combination Product (y/n)N
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number186715650
Device Catalogue Number186715650
Device Lot Number207036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2019
Date Manufacturer Received11/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-