LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-01-04
Device Problem Pumping Problem (3016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/11/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).A livanova service representative was dispatched to the hospital and confirmed that the cp5 drive head appeared to have water damages.The affected cp5 drive unit was requested back to livanova (b)(4) for investigation.During the internal visual inspection it was confirmed that several components were affected by dried fluid residuals and oxidation, which made any electrical test impossible.As root cause fluid/moisture ingress was determined, due to an improper handling by the customer.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
 
Event Description
Livanova (b)(4) received a report that whilst practicing an oxygenator change-out during a training session, a s5 system alarmed showing a centrifugal pump 5 (cp5) error message.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9174333
MDR Text Key194491820
Report Number9611109-2019-00806
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 10/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-04
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-