This is being filed to report a leak.It was reported that during preparation of a steerable guide catheter (sgc), the sgc was flushed; however, an air bubble was observed in the guide chamber.The sgc was re-flushed and a small drop of water was observed in the fluid column which indicated a possible leak.A decision was made not use the sgc in the patient.The procedure was successfully completed with a new sgc.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
|
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.Based on the returned device analysis the reported leak could not be confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents reported from this lot.All available information was investigated and a definitive cause for the reported leak in this incident could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
|