• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SURESHOT TARGETER; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SURESHOT TARGETER; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 71692801
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/18/2019
Event Type  malfunction  
Event Description
It was reported that during surgery, the information on the monitor showed the following message: "the targeter is broken, please exchange".A c- arm was used to complete the surgery.  no delay.No impact or injury to patient reported.
 
Manufacturer Narrative
It was reported that during surgery the sureshot displayed a message that indicated that the targeter was broken or non-valid.The associated sureshot targeter was returned and evaluated.Visual inspection of the returned product found blemishes and stains on the device.The review of the manufacturing records for the listed serial number did not reveal any deviation from the standard manufacturing processes.A functional evaluation was performed by connecting the sureshot targeter to the sureshot interface unit.An error message was displayed which read, ¿sureshot targeter invalid or broken tool - please exchange.¿ thus, the stated failure was confirmed.The targeter is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Damage from repeated use can occur and some causes for the malfunction would include a broken cable, damaged connector, broken srom connection, and/or silicone overmolding failure.Our investigation found that the subject device has been previously evaluated for similar events and an upgrade to the targeter has taken place.A second generation targeter has been released and is available (please reference device 71692851) for use.The sureshot device user manual is available, identifying pre-operative requirements for use and troubleshooting suggestions if needed.No additional actions are being taken at this time; however smith and nephew will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESHOT TARGETER
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9187125
MDR Text Key162443135
Report Number1020279-2019-03637
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K092497
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 11/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71692801
Device Lot NumberND5074
Date Manufacturer Received11/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age30 YR
Patient Weight65
-
-