CAREFUSION ALARIS® PUMP MODULE ADMINISTRATION SET; SET,ADMINISTRATION,INTRAVASCULAR Back to Search Results
Model Number 2420-0007
Device Problem Stretched (1601)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/17/2019
Event Type  malfunction  
Manufacturer Narrative
The customer¿s report that a bubble was discovered in the silicone tubing was confirmed.During inspection it was observed that the silicone segment tubing had a bubble starting at the ring retainer of the upper fitment.The set was visually inspected for kinks, incomplete bonding engagements, holes/ tears in the tubing or damages to the components.During inspection it was observed that the silicone segment tubing had a bubble.No other anomalies were observed on the set during visual inspection.Clear liquid was observed inside both the set¿s tubing and the drip chamber.Functional testing was performed and showed no anomalies.The balloon is caused by excessive pressure within the silicone segment.The root cause for the source of the excessive pressure was not definitively determined.
 
Event Description
It was reported that after the rn noted the pump alarming, and upon checking the tubing, it was noticed the set had a "balloon" in the pumping segment.It was also confirmed that there was no patient harm as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS® PUMP MODULE ADMINISTRATION SET
Type of Device
SET,ADMINISTRATION,INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
mustafa omar
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key9187673
MDR Text Key162877645
Report Number9616066-2019-02843
Device Sequence Number1
Product Code FPA
UDI-Device Identifier07613203021012
UDI-Public7613203021012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2420-0007
Device Catalogue Number2420-0007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-