Catalog Number 114827
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/30/2019
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as loosening. The previous surgery and the surgery detailed in this event occurred 9 days apart. Initial and prolonged hospitalization is required. The healthcare professional indicated there was no delay in surgery and another suitable device was available for use. The revision surgery was completed as intended. The devices were disposed of at the hospital and not made available to djo surgical for examination. A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements. There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event. The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery. Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review. The root cause of this complaint was a revision surgery due to loosening. There were no findings during this evaluation that indicate the reported devices were defective. No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event. There are multiple factors that may contribute to an event that are outside the control of djo surgical. There are no indications of a product or process issue affecting implant safety or effectiveness. Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event Description
Revision surgery - due to humeral became loose.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDISCOVERY ELBOW
Type of DeviceDISC ULNA 4X115MM RT W BRNG C
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key9193065
MDR Text Key162451149
Report Number1644408-2019-00965
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Expiration Date11/22/2021
Device Catalogue Number114827
Device Lot Number804050
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/17/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/15/2019 Patient Sequence Number: 1
-
-