The reported oad was received at csi for analysis.
Visual examination did not reveal any damage.
The device was turned on and off numerous times with no issues observed.
While performing function testing, the motor cable set, start switch and potted pcb were soaked with saline solution and the power button was activated with no unintended spinning observed.
The oad was turned off and rested without the oad turning back on.
There were no visible leak paths, and no damage to the start switch cables or the cable retainer.
Further examination revealed there was no damage to the motor, cable set, brake or speed switch ribbons and no damage observed that would have contributed to the reported event.
At the conclusion of the device analysis, the reported event that the oad spun independently without being prompted could not be confirmed as the oad functioned as intended during analysis.
The device history record for this oad lot number has been reviewed.
No issues or discrepancies were noted during this review that would have contributed to the reported event.
The device met material, assembly, and quality control requirements.
Csi id# (b)(4).
|
The peripheral diamondback orbital atherectomy device (oad) spun independently without being prompted.
During testing outside of the body, the oad intermittently started and stopped spinning and the saline infusion pump was used to troubleshoot the issue.
After 2 to 3 seconds the oad spun without being prompted at 60,000 rpms for approximately 2 to 3 seconds before becoming caught on sterile towels in the field.
The device was replaced with a second oad to complete the procedure with no major delays.
The patient was discharged without any complications.
|