MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problems Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 8709, lot#: j0058223r, implanted: (b)(6) 2001, product type: catheter. Other relevant device(s) are: product id: 8709, serial/lot #: (b)(4), ubd: 2003-03-06, udi#: (b)(4). If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider (hcp) regarding a patient who was receiving an unknown drug of unknown concentration at an unknown dose rate via an implantable pump for intractable spasticity, cerebral palsy, and other. It was reported that the patient was needing a brain mri (magnetic resonance imaging). The patient had an abdominal x-ray performed on (b)(6) 2019 and it was discovered that a fragment of the catheter had broken off and was in the spinal canal. The date of the event was unknown. The hcp was questioning how this may affect mri compatibility. No patient symptom was reported. No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key9198678
MDR Text Key170268440
Report Number3004209178-2019-19765
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Type of Report Initial
Report Date 10/16/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/16/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/14/2017
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/03/2019
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured07/21/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-