Model Number EVOLUTR-34-US
Device Problems Separation Failure (2547); Malposition of Device (2616); Insufficient Information (3190)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/24/2019
Event Type  Injury  
Manufacturer Narrative
Product analysis: the product was not returned; therefore, no product analysis can be performed.Conclusion: without return of the product, no definitive conclusions could be drawn regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, a second valve was implanted for an unknown reason.It was reported the first valve was left implanted in the descending aorta.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Concomitant medical products: information references the main component of the system.Other relevant device(s) are: product id: envpro-16-us.Additional information was received that the failure mode of the first valve was that the valve could not be recaptured a second time.The valve was pulled down into the abdominal aorta and deployed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic received additional information that identified this as a duplicate event previous submitted in regulatory report #2025587 -2019-02979.The serial number of the valve was corrected in section d4.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT R TRANSCATHETER AORTIC VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key9210838
MDR Text Key163040214
Report Number2025587-2019-03198
Device Sequence Number1
Product Code NPT
UDI-Device Identifier00763000017682
UDI-Public00763000017682
Combination Product (y/n)N
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/04/2020
Device Model NumberEVOLUTR-34-US
Device Catalogue NumberEVOLUTR-34-US
Was Device Available for Evaluation? No
Date Manufacturer Received10/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age85 YR
Patient Weight128
-
-