• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER Back to Search Results
Catalog Number 7209485
Device Problem Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/24/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported that during the procedure, the loop retriever was found to be already disassembled between head and shaft and it was noticed when the package was opened.There was no delay or patient injuries but it is unknown how the procedure was finished since no backup device was available to complete it.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
One 7209485 disposable meniscus mender ii set reported on.The complaint stated: ¿it was noticed that there are issues fitting the accupass monofilament down the meniscus mender cannulation.The problem started happening regularly since a few months, as the monofilament stopped working the way it used to.¿ the product was not returned for evaluation.Due to product unavailability, the complaint could not be ultimately confirmed.Definitive conclusions, accurate investigation and evaluation were not fully possible without evaluation of physical product.If objective evidence, relevant information, packaging or product becomes available to assist with evaluation, the complaint will certainly be revisited.Factors that may affect device performance include: device ability, surgical ability, implant location and tissue condition.Review of instruction for use documentation confirms instructions, precautionary statements and recommendations for proper use of product.Engineering evaluation confirmed the product met current specifications at the time of distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET MENISCUS MENDER II DISPOSABLE
Type of Device
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key9216666
MDR Text Key163133178
Report Number1219602-2019-01317
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier03596010468482
UDI-Public03596010468482
Combination Product (y/n)N
PMA/PMN Number
K885311
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7209485
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received11/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-