• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER; HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER; HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES) Back to Search Results
Model Number G17147
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/08/2019
Event Type  malfunction  
Manufacturer Narrative
A follow up report will be submitted should additional relevant information become available.
 
Event Description
It was reported, during an unknown procedure in which a cook silicone balloon hysterosalpingography injection catheter was used, that prior to patient contact, when filling the balloon it ruptured.It is unknown how the procedure was completed.It was reported that the patient did not experience any adverse effects due to this occurrence.Additional information has been requested regarding the patient and the event.At the time of this report, no further information has been provided.
 
Event Description
No new patient or event information since the last report was submitted on 22oct2019.
 
Manufacturer Narrative
H6 ec method code desc - 5: communication/interview (4111).Investigation/evaluation: a visual inspection of the returned device packaging was conducted.A document-based investigation was also performed including a review of complaint history, device history record, the instructions for use, quality control data, and specifications.One open package labeled cook silicone balloon hysterosalpingography injection catheter was returned for investigation.Visual examination confirmed the package was received empty.No product was inside of the packaging or the shipping box.The complaint device was not returned; therefore, no physical examinations could be performed; however, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other complaints associated with the complaint device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other related complaints from the lot have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: warning: always inflate the balloon with sterile liquid.Never inflate with air, carbon dioxide or any other gas.Precautions: do not over inflate.Using excessive pressure to inflate the balloon on this device can cause the balloon to rupture.Refer to product label or the inflation check valve on the balloon device for appropriate balloon volume.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.Balloons are leak tested using air during manufacturing and again during inspection processes.The complaint was confirmed based on customer testimony.A definitive cause of the event could not be determined.Per the quality engineering risk assessment, no further action is warranted.Cook medical will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER
Type of Device
HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES)
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9218559
MDR Text Key163169821
Report Number1820334-2019-02670
Device Sequence Number1
Product Code HES
UDI-Device Identifier00827002171473
UDI-Public(01)00827002171473(17)211227(10)9407656
Combination Product (y/n)N
PMA/PMN Number
K891290
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,other,use
Type of Report Initial,Followup
Report Date 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/27/2021
Device Model NumberG17147
Device Catalogue NumberJ-CHSG-503000
Device Lot Number9407656
Was Device Available for Evaluation? No
Date Manufacturer Received11/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-