The patient was undergoing a thrombectomy procedure in the common carotid artery (cca)/right internal carotid artery (ica) using a penumbra system aspiration pump max 220 (pump max).During the procedure, the pump max started emitting smoke and stopped working.The physician was unable to turn the pump max back on and there was no aspiration.Therefore, the pump max was disconnected and the procedure was completed using a manual aspiration with a syringe.There was no report of an adverse effect to the patient.
|
Results: the fuse port on the back of the pump was elevated.The fuse port was removed, and corrosion was found on the fuses.The pump max housing was opened, and blood was observed within the pump housing.Liquid was within the tubing and the piston.During functional testing, the pump max was powered on and the fan could be heard but no aspiration was observed.Conclusions: evaluation of the returned pump max revealed liquid within the tubing and pistons.This typically occurs if the tubing is connected directly to the pump instead of the canister, supplied by penumbra.If liquid is aspirated into the pump, the pump may not function properly.Further evaluation revealed corrosion within the pump housing and on the fuses on the back of the pump housing.These damages were likely a result of liquid being aspirated into the pump.Penumbra pumps are 100% visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|