Model Number SUPPORT ARM 178
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/08/2019
Event Type  malfunction  
Event Description
It was reported that the support arm for patient tubes, mounted on the ventilator carrier, broke.There was no patient harm.(b)(4).
 
Event Description
Manufacturer's ref #: (b)(4).
 
Manufacturer Narrative
The support arm was not returned for investigation.According to provided pictures and additional information, the ball bearings on the rail clamp mounted on the support arm were stuck.This has most likely contributed to an inability to fasten the support arm on the side rail mounted on the ventilator carrier.The root cause of the stuck ball bearings has not been determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPPORT ARM 178
Type of Device
SUPPORT, ARM
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
MDR Report Key9223759
MDR Text Key174327671
Report Number8010042-2019-00776
Device Sequence Number1
Product Code IOY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2019
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSUPPORT ARM 178
Device Catalogue Number6693715
Was Device Available for Evaluation? No
Date Manufacturer Received10/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-