PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC BASE ANALYZER, PACEMAKER GENERATOR FUNCTION Back to Search Results
Model Number 24970A
Device Problems Display or Visual Feedback Problem (1184); Pacing Problem (1439)
Patient Problem Bradycardia (1751)
Event Date 07/09/2019
Event Type  Injury  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that the programmer displayed a certificate renewal request while a patient was being paced with the analyzer. The patient experienced a heart rate of 30 beats per minute (bpm). The user was unable to update the security certificate at the time because no internet was available in the operating room. A different model of programmer was used to pace the patient. The programmer remains in use. No further patient complications were noted.

 
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARELINK SMARTSYNC BASE
Type of DeviceANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9234661
MDR Text Key163712510
Report Number3004593495-2019-01119
Device Sequence Number1
Product Code DTC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberP890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/02/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received10/24/2019
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number24970A
Device Catalogue Number24970A
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/26/2019
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient TREATMENT DATA
Date Received: 10/24/2019 Patient Sequence Number: 1
-
-