Model Number 186-0106 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/24/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, the device could not provide correct readings.The customer indicated no patient harm.
|
|
Manufacturer Narrative
|
Evaluation summary: one device was inspected.The device failed to meet specification as it was received or made available for evaluation.The device was visually inspected and no physical anomalies were found.The device was tested with a sensor lot code tester; the device was within expiration date and displayed a ¿sensor valid¿ message.An impedance check was performed using a digital multimeter, and all electrodes passed the impedance check.The investigation found the device to function normally.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
|