(b)(4).Device lot number: 2100809308, d.O.M: (b)(6) 2019.Device lot number: 2100800481, d.O.M: (b)(6) 2019.Device lot number: 2100801533, d.O.M: (b)(6) 2019.The complaint devices rt265 infant dual heated evaqua2 breathing circuits are currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
|
(b)(4).Device 1, 5 and 6: lot unknown.Device 4 lot number: 2100809308, d.O.M: 28 june 2019.Device 2 lot number: 2100800481, d.O.M: 19 june 2019.Device 3 lot number: 2100835013, d.O.M: 25 july 2019.Method: the swivels of complaint rt265 infant dual heated evaqua2 breathing circuits were returned to fisher & paykel healthcare (f&p) in new zealand where they were visually inspected and pressure tested.Results: device 1: the swivel elbow and swivel wye were returned partly disassembled.Damages was observed to the swivel wye.The swivel elbow and swivel wye were reassembled.The pressure test for the complaint device confirmed a tight fit of swivel components.Device 2 & 4: the swivel elbow and swivel wye were returned assembled.No damages were observed to any of the swivel components.The pressure test for the complaint device confirmed a tight fit of swivel components.Device 3, 5, 6: the swivel elbow and swivel wye were returned assembled.No damages were observed to any of the swivel components.The pressure test for the complaint device confirmed a loose fit of swivel components.Conclusion: investigation into this complaint reviewed the manufacturing process (operator, equipment, measurement and environment), process documentation, samples of product, complaint devices and performed a material analysis.The investigation indicated a potential resin material mix-up was the most likely cause.We have since implemented an additional material verification step, at the point of resin material addition to the moulding machine feed.This verification would identify any potential resin material mix-up prior to use.All rt265 dual-heated evaqua2 breathing circuits are designed to conform to iso:5367.All rt265 circuits are visually inspected, pressure and flow tested during production and those that fail, are rejected.The user instructions that accompany the rt265 infant dual heated evaqua2 breathing circuit also state the following: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "set appropriate ventilator alarms.".
|