Model Number 60-02-60 |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/04/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.He replaced the processor board and the issue was solved.Functional verification testing was completed without further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The pump serial read-out has been provided.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a centrifugal pump 5 (cp5) stopped functioning with no flow during procedure.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H.10: the reported malfunction could be confirmed by the analysis of the serial read-out of the hkr board.The replaced board was requested by livanova for a deeper investigation.Results revealed that the root cause of the reported pump stop was a deviation in the internal memory bank of the microcontroller.
|
|
Event Description
|
See initial report.
|
|