• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-60
Device Problem Pumping Stopped (1503)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/04/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.He replaced the processor board and the issue was solved.Functional verification testing was completed without further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The pump serial read-out has been provided.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump 5 (cp5) stopped functioning with no flow during procedure.There was no report of patient injury.
 
Manufacturer Narrative
H.10: the reported malfunction could be confirmed by the analysis of the serial read-out of the hkr board.The replaced board was requested by livanova for a deeper investigation.Results revealed that the root cause of the reported pump stop was a deviation in the internal memory bank of the microcontroller.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key9255552
MDR Text Key195731823
Report Number9611109-2019-00859
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-60
Was the Report Sent to FDA? No
Date Manufacturer Received01/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-