ENCORE MEDICAL L.P. MOVATION KNEE SYSTEM MOVATION TIBIAL INSERT, SIZE 8, 9MM VE Back to Search Results
Catalog Number 292-09-708
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Unspecified Infection (1930)
Event Date 09/30/2019
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as an infection. The healthcare professional indicated there was no delay in surgery and another suitable device was available for use. The revision surgery was completed as intended. The device was disposed of at the hospital and not made available to djo surgical for examination. A review of the device history record(s) show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements. There were no nonconforming material reports associated with the product that may have contributed to the reported event. The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery. Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review. The root cause of this complaint was a revision surgery due to an infection. There were no findings during this evaluation that indicate that the reported device was the source or had a direct connection with the patient's infection. No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system. There are multiple factors that may contribute to an infection that are outside of the control of djo surgical. There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event Description
Revision surgery - due to an infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOVATION KNEE SYSTEM
Type of DeviceMOVATION TIBIAL INSERT, SIZE 8, 9MM VE
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, TX 78758-5445
MDR Report Key9257416
MDR Text Key164439490
Report Number1644408-2019-01055
Device Sequence Number1
Product Code OIY
UDI-Device Identifier00888912112437
UDI-Public00888912112437
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121727
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Expiration Date10/28/2019
Device Catalogue Number292-09-708
Device Lot Number328N1007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/30/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/16/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/30/2019 Patient Sequence Number: 1
-
-