PLEXUS MANUFACTURING SDN. BHD CARELINK ENCORE PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 29901A
Device Problems Computer Software Problem (1112); Failure to Power Up (1476); Battery Problem (2885); Computer Operating System Problem (2898)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer wouldn't turn on and had no lights.It had trouble booting up multiple times on various days.It was noted to sometimes get stuck on one screen and at other times an error would pop up stating there was a problem.It was also noted that there had been errors stating that the power was not sufficient when the battery was testing ok.This was surprising to the user as the programmer was always plugged in.At one point it was discovered that the programmer was plugged into an outlet that had no power.The programmer was plugged into a different outlet and powered on as expected.The issue later persisted.The user then attempted pulling the battery out and unplugging the programmer to reboot, which did not resolve the issue.The programmer was returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis:analysis confirmed that the programmer battery gave a message to replace the battery.The battery tested to be ok.Replaced the battery, reloaded and updated software.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK ENCORE PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9257467
MDR Text Key175244716
Report Number3004593495-2019-01139
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00643169763548
UDI-Public00643169763548
Combination Product (y/n)N
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number29901A
Device Catalogue Number29901A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2019
Date Manufacturer Received10/31/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-