Model Number 29901A |
Device Problems
Computer Software Problem (1112); Failure to Power Up (1476); Battery Problem (2885); Computer Operating System Problem (2898)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/08/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the programmer wouldn't turn on and had no lights.It had trouble booting up multiple times on various days.It was noted to sometimes get stuck on one screen and at other times an error would pop up stating there was a problem.It was also noted that there had been errors stating that the power was not sufficient when the battery was testing ok.This was surprising to the user as the programmer was always plugged in.At one point it was discovered that the programmer was plugged into an outlet that had no power.The programmer was plugged into a different outlet and powered on as expected.The issue later persisted.The user then attempted pulling the battery out and unplugging the programmer to reboot, which did not resolve the issue.The programmer was returned for service.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product analysis:analysis confirmed that the programmer battery gave a message to replace the battery.The battery tested to be ok.Replaced the battery, reloaded and updated software.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
|